Advanced Instruments Inc.
2855 Metropolitan Place, Pomona, California 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, e-mail: [email protected]
GPR-1600 MS
ppm Oxygen Analyzer
Owner’s Manual
Vue de la page 0
1 2 3 4 5 6 ... 49 50

Résumé du contenu

Page 1 - Owner’s Manual

Advanced Instruments Inc. 2855 Metropolitan Place, Pomona, California 91767 USA ♦ Tel: 909-392-6900, Fax: 909-392-3665, e-mail: [email protected]

Page 2 - 1 Introduction

Advanced Instruments Inc. 10 Oxygen, the fuel for this electrochemical transducer, reacts chemically at the sensing electrode to produce an

Page 3

Advanced Instruments Inc. 11 Calibration & Accuracy Overview Single Point Calibration: As previously described the galvanic oxygen

Page 4 - Oxygen Service

Advanced Instruments Inc. 12 Zero Calibration In theory, the galvanic fuel cell type oxygen has an absolute zero meaning it produces no sig

Page 5 - 3 Safety Guidelines

Advanced Instruments Inc. 13 * Refer to analyzer specifications for comparable data on the Pico-Ion UHP and MS oxygen sensors.

Page 6

Advanced Instruments Inc. 14 3. Note: The proximity of the analyzer to the sample point and use of optional sample conditioning components

Page 7

Advanced Instruments Inc. 15 Mounting GPR-1600MS-W Option:

Page 8

Advanced Instruments Inc. 16 Mounting GPR-1600MS-W-306 Option:

Page 9 - 5 Operation

Advanced Instruments Inc. 17 Gas Connections The GPR-1600MS with its standard flow through configuration is designed for positive pressure s

Page 10 - Advanced Instruments Inc

Advanced Instruments Inc. 18 Electrical Connections The appropriate AC power requirement must specified be specified at order placement if t

Page 11

Advanced Instruments Inc. 19 Procedure: 1. As illustrated above the sensor, power and alarm relays and signal output connections are hard

Page 12

Advanced Instruments Inc. 2 Table of Contents Introduction 1 Quality Control Certification 2 Safety & Installation 3 Features

Page 13

Advanced Instruments Inc. 20 Alarms Alarm 1 and Alarm 2 represent two threshold type alarms that can be configured in the field from the ana

Page 14

Advanced Instruments Inc. 21 Temperature Controlled Heater System with Runaway Protection Circuit The standard GPR-1600MS Series analyzer is

Page 15

Advanced Instruments Inc. 22 2. Purge the oxygen trapped in the newly connected gas lines for 3-5 minutes. 3. Flow zero gas or sample gas

Page 16

Advanced Instruments Inc. 23 START-UP TEST ELECTRONICS – PASS TEMP SENSOR – PASS BAROMETRIC SENSOR – PASS REV. 1.61

Page 17

Advanced Instruments Inc. 24 Note: For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the fu

Page 18

Advanced Instruments Inc. 25 8. The display will not shift automatically. Instead, when the oxygen reading (actually the sensor’s signal ou

Page 19

Advanced Instruments Inc. 26 GAS CONCENTRATION PERCENT PPM >>> 01.0 SET ALARM 1 VALUE PRESS UP OR DOWN TO CHANGE VALUE E

Page 20

Advanced Instruments Inc. 27 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATION CONFIG ALARMS BYPASS ALARMS >>> MAIN MENU SET A

Page 21

Advanced Instruments Inc. 28 3.3PPM AUTO SAMPLING 10 PPM RANGE 24.5 C LO1 2 PPM 10 PPM HI2 7

Page 22

Advanced Instruments Inc. 29 4. Press the ENTER key to bypass and disable both the local audible alarm and relay contacts. After 3 seconds,

Page 23 - 3.3PPM

Advanced Instruments Inc. 3 2 Quality Control Certification Date: Customer:

Page 24

Advanced Instruments Inc. 30 recommended only when performing a ZERO CALIBRATION and during troubleshooting and should not be repeated befor

Page 25

Advanced Instruments Inc. 31 CALIBRATION SPAN CALIBRATE ZERO CALIBRATE DEFAULT SPAN DEFAULT ZERO OUTPUT SPAN OUTPUT ZERO 11. Advance th

Page 26

Advanced Instruments Inc. 32 Default Zero: The software will eliminate any previous zero calibration adjustment and display the actual the

Page 27

Advanced Instruments Inc. 33 9. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cu

Page 28

Advanced Instruments Inc. 34 routine subsequent SPAN CALIBRATION. ¾ If a ZERO CALIBRATION adjustment has NOT been made as described above,

Page 29

Advanced Instruments Inc. 35 CALIBRATION SPAN CALIBRATE ZERO CALIBRATE DEFAULT SPAN DEFAULT ZERO OUTPUT SPAN OUTPUT ZERO 7. Advance

Page 30 - 3.3PPM

Advanced Instruments Inc. 36 15. Regulate the pressure and control the flow rate as described above between 5-30 psig and a 2 SCFH flow rat

Page 31

Advanced Instruments Inc. 37 PASSED CALIBRATION OR FAILED CALIBRATION 33. If the calibration is successful, the analyzer returns t

Page 32

Advanced Instruments Inc. 38 099.0 OUTPUT SPAN OFFSET PRESS UP OR DOWN TO CHANGE VALUE ENTER TO SAVE MENU TO RETURN 9. Press th

Page 33

Advanced Instruments Inc. 39 Sampling GPR-1600MS Oxygen Analyzer requires positive pressure to flow the sample gas by the sensor to measure

Page 34

Advanced Instruments Inc. 4 Certificate of Cleaning Oxygen Service Standard: Manufacturing Procedure No. P-1057 Rev-1, Compressed Ga

Page 35

Advanced Instruments Inc. 40 6 Maintenance There are no moving parts in the analyzer given the modular nature of the electronics and senso

Page 36

Advanced Instruments Inc. 41 7 Spare Parts Recommended spare parts for the GPR-1600MS Oxygen Analyzer include: Item No. Description GP

Page 37

Advanced Instruments Inc. 42 8 Troubleshooting Symptom Possible Cause Recommended Action Slow recovery At installation, def

Page 38

Advanced Instruments Inc. 43 Symptom Possible Cause Recommended Action Erratic O2 reading or No O2 reading Change in

Page 39

Advanced Instruments Inc. 44

Page 40 - 6 Maintenance

Advanced Instruments Inc. 45 9 Warranty The design and manufacture of GPR Series oxygen analyzers, monitors and oxygen sensors are perform

Page 41 - 7 Spare Parts

Advanced Instruments Inc. 46 10 MSDS Material Safety Data Sheet Product Identification Product Name Oxygen Sensor Series - PSR, GPR, AII,

Page 42 - 8 Troubleshooting

Advanced Instruments Inc. 47 Spill or Leak Steps if material is released Sensor is packaged in a sealed plastic bag, check the sensor ins

Page 43

Advanced Instruments Inc. 48 Appendix B Correlating Readings - LCD Display and 4-20mA Output In rare instances the 4-20mA signal output ma

Page 44

Advanced Instruments Inc. 49 100.0 OUTPUT ZERO OFFSET PRESS UP OR DOWN TO CHANGE VALUE ENTER TO SAVE MENU TO RETURN 14. Enter

Page 45 - 9 Warranty

Advanced Instruments Inc. 5 3 Safety Guidelines Safety This section summarizes the basic precautions applicable to all analyzers. Additio

Page 46

Advanced Instruments Inc. 50 27. Enter the calculated adjustment value, refer to example described above. Note: Once the initial adjustment

Page 47

Advanced Instruments Inc. 6 Installation Gas Sample Stream: Ensure the gas stream composition of the application is consistent with the spe

Page 48 - Appendix B

Advanced Instruments Inc. 7 Inlet Pressure: Analyzers designed for flowing samples under positive pressure or pump vacuum (for samples at a

Page 49

Advanced Instruments Inc. 8 4 Features & Specifications

Page 50

Advanced Instruments Inc. 9 5 Operation Principle of Operation The GPR-1600MS ppm Oxygen Analyzers incorporates a proprietary Pico-Ion oxy