Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 3 Sample System: The GPR-3100 is supplied with sta
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 4 throttle valve on the pump’s inlet side may be n
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 5 Calibration and Accuracy Single Point Calibratio
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 6 Graph C illustrates these 'worse case'
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 7 Installation The GPR-3100 Oxygen Analyzer consis
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 8 Mounting the Analyzer: The GPR-3100’s 10.8”W x 7
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 9 Gas Connections: The GPR-3100 is designed for po
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 10 Procedure: Caution: Do not change the factory s
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 11 Output Connections As illustrated above the sen
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 12 Power Failure Alarm A dry contact rated at 30VD
Advanced Instruments Inc. Table of Contents Introduction 1 Quality Control Certification 2 Safety 3 Features & Specifications 4 Operation
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 13 Installing the Oxygen Sensor The GPR-3100 is eq
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 14 Installing Span Gas Caution: Do not contaminat
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 15 Establishing Power to the Electronics: Once the
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 16 Operating the Menu Driven Controls – The appen
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 17 Range Selection: The analyzer display defaults
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 18 4. Press the ENTER key when F-C starts flashin
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 19 Calibration - General Guideline Zeroing the ana
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 20 consume the oxygen that has dissolved into the
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 21 span calibration again for up to 3 months under
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 22 16. The analyzer returns to the SAMPLE mode aft
Advanced Instruments Inc. 1 Introduction Thank You For Your Business Your new oxygen analyzer is a precision piece of equipment designed to g
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 23 To avoid damaging the sensor or sample system o
Advanced Instruments Inc. 6 Maintenance GPR-3100 Rev 7/2004 6
Advanced Instruments Inc. 7 Spare Parts Recommended spare parts for the GPR-3100 % High Purity O2 Analyzer include: Item No. DescriptionGPR-11
Advanced Instruments Inc. 8 Troubleshooting Symptom Possible Cause Recommended Action Reading does not reflect expected values Sensor was not c
Advanced Instruments Inc. 9 Warranty What is covered: Any defect in material and workmanship from normal use in accordance with the Owner’s Manu
Advanced Instruments Inc. 9 Warranty What is not covered: This warranty does not cover installation; defects resulting from accidents; damage wh
Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Product Identification Product Name Oxygen Sensor Models CAD, GPR, PSR, SAF, 670
Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Fire and Explosion Data Flash and Fire Points Not applicable Flammable Limits N
Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Symptoms Eye contact - burning sensation. Skin contact - soapy slick feeling. Me
Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t
Advanced Instruments Inc. 2 Quality Control Certification Date: Customer: Order No.: Pass Mo
Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t
Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t
Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t
Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t
Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t
Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t
Advanced Instruments Inc. 3 Safety This section summarizes the essential generic precautions applicable to all analyzers. Additional precautions
Advanced Instruments Inc. 3 Safety Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the operato
Technical Specifications Accuracy: < 1% of FS range at constant conditions; inherent accu-racy of 0.1% following calibration with 95-100% oxygen
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 1 Principle of Operation The GPR-3100 oxygen analy
Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 2 With the sensor and sample gas lines temperature