Advanced Instruments Inc. 10 Zero Calibration In theory, the galvanic fuel cell type oxygen has an absolute zero meaning it produces no signal o
Advanced Instruments Inc. 11 Span Calibration Span Calibration involves adjusting the transmitter electronics to the sensor’s signal output at a
Advanced Instruments Inc. 12 ¾ When certified span gas is not available to calibrate a ppm analyzer (immediately following air calibration reintro
Advanced Instruments Inc. 13 Installing the Oxygen Sensor The GPR-3500MO portable oxygen purity analyzer is equipped with an integral oxygen senso
Advanced Instruments Inc. 14 Span Gas Preparation Caution: Do not contaminate the span gas cylinder when connecting the regulator. Bleed the air
Advanced Instruments Inc. 15 Charging the battery requires a common 9VDC adapter (positive pole located inside the female connector) supplied with
Advanced Instruments Inc. 16 Zero Calibration In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. Ho
Advanced Instruments Inc. 17 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATE 24.5 C 100 KPA >>> CAL
Advanced Instruments Inc. 18 Output Zero Accuracy due to manufacturer tolerances may result in a slight difference between the LCD display and the
Advanced Instruments Inc. 19 Span Calibration Maximum drift from calibration temperature is approximately 0.11% of reading per °C. The analyzer ha
Advanced Instruments Inc. 2 Table of Contents Introduction 1 Quality Control Certification 2 General Safety & Installation 3 Features &
Advanced Instruments Inc. 20 Caution: To assure an accurate calibration, the temperature and pressure of the span gas must closely approximate the
Advanced Instruments Inc. 21 100.00 % SPAN CALIBRATION ENTER TO CALIBRATE MENU TO ABORT 23. Press ENTER to initiate the span calibration pr
Advanced Instruments Inc. 22 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATION CONFIG ALARMS BYPASS ALARMS >>> CALIBRATION SPAN CALIBR
Advanced Instruments Inc. 23 Sampling GPR-3500MO Oxygen Portable oxygen purity analyzer requires positive pressure to flow the sample gas by the s
Advanced Instruments Inc. 24 6 Maintenance With exception of components related to optional equipment and charging the battery of portable analyze
Advanced Instruments Inc. 25 Battery Charging the battery requires a common 9VDC adapter (positive pole located inside the female connector) suppl
Advanced Instruments Inc. 26 8 Troubleshooting Symptom Possible Cause Recommended Action Reading does not reflect expected values Sensor was
Advanced Instruments Inc. 27 9 Warranty The design and manufacture of GPR Series oxygen analyzers, monitors and oxygen sensors are performed und
Advanced Instruments Inc. 28 10 MSDS – Material Safety Data Sheet Product Identification Product Name Oxygen Sensor Series - PSR, GPR, AII, XLT
Advanced Instruments Inc. 29 Reactivity Data Stability Stable Conditions Contributing to Instability None Incompatibility KOH = Avoid contact
Advanced Instruments Inc. 3 Advanced Instruments Inc. appreciates your business and pledges to make every effort to maintain the highest possible q
Advanced Instruments Inc. 30
Advanced Instruments Inc. 31 Output Zero Procedure: 1. Access the MAIN MENU by pressing the MENU key. 2. Advance the reverse shade cursor using t
Advanced Instruments Inc. 32 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATION CONFIG ALARMS BYPASS ALARMS >>> CALIBRATION SPAN CALIBRA
Advanced Instruments Inc. 4 3 General Safety & Installation Safety This section summarizes the basic precautions applicable to all analy
Advanced Instruments Inc. 5 Installation Gas Sample Stream: Ensure the gas stream composition of the application is consistent with the specific
Advanced Instruments Inc. 6 Flow rates of 1-5 SCFH cause no appreciable change in the oxygen reading. However, flow rates above 5 SCFH generate bac
Advanced Instruments Inc. 7 Moisture and/or particulates generally can be removed from the sensor by opening the sensor housing and either blowing
Advanced Instruments Inc. 8 The sensor employed by the GPR-3500MO is but one part of the proprietary advancements in sensor technology. It is the o
Advanced Instruments Inc. 9 Accuracy & Calibration Single Point Calibration: As previously described the galvanic oxygen sensor generates