Advanced Instrtuments Inc.
5 Operation - Appendix A
* MAIN MENU Sample
Sample
Span
Zero
Alarm
System
Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48
99.55 %
MAIN MENU Sample SPAN Sample * MAIN MENU Sample
Sample * Factory Default (1) Sample (1) SPAN menu, Factory Default option -
* Span Calibrate Span Purpose is to reset the microprocessor as the first step in
Zero Zero troubleshooting perceived abnormalities. The span value
Alarm Alarm (which may be compromised) is reset using an average
System System sensor output value to temporarily assess the analyzer's
performance. Calibration is required upon completion.
Auto Range 90 to 100 % Auto Range 90 to 100 % Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48 75F 101 Kpa 02/10/2004 17:32:48 75F 101 Kpa 02/10/2004 17:32:48
Move cursor (*) to Span using DOWN ARROW key Move cursor (*) to Factory Default using DOWN ARROW key
Press ENTER to select Press ENTER to select
Activates (1) and returns to MAIN MENU
Note: Returns to MAIN MENU if no entry within 30 seconds
SPAN Sample Span Gas Sample 100.00 % Sample * MAIN MENU Sample
Factory Default Enter as PPB * Sample
* Calibrate Enter as PPM Press Up or Down Span
* Enter as % to change value (2) Zero
Enter to Save Alarm
ESC to Return System
Auto Range 90 to 100 % Auto Range 90 to 100 % Auto Range 90 to 100 % Auto Range 90 to 100 %
75F 101 Kpa 02/10/2004 17:32:48 75F 101 Kpa 02/10/2004 17:32:48 75F 101 Kpa 02/10/2004 17:32:48 75F 101 Kpa 02/10/2004 17:32:48
Move cursor (*) to Calibrate using DOWN ARROW key Move cursor (*) to desired Man Range using DOWN ARROW key (2) To change value of 1st digit.
Press ENTER to select Press ENTER to select Press ENTER to advance cursor (*) to 2nd digit
Returns to MAIN MENU Press Up or Down to change value
Repeat as necessary for remaining digits
99.55 % 99.55 %
99.55 %
5_ops_3100_menu 11/22/2005 Page 2 of 7