Analytical-industries GPR-1200 MS ATEX Portable PPB Oxygen Analyzer Manuel d'utilisateur Page 27

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 32
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 26
27
Symptom Possible Cause Recommended Action
Erratic O
2
reading
or
No O
2
reading
Test sensor independent from analyzer
Change in sample pressure
Dirty electrical contacts in upper section of
sensor housing
Corroded solder joints on sensor PCB from
corrosive sample or electrolyte leakage from
sensor
Corroded spring loaded contact in upper
section of sensor housing from liquid in
sample or electrolyte leakage from sensor
Liquid covering sensing area
Improper sensor selection
Presence of interference gases
Presence of sulfur gases
Unauthorized maintenance
Sensor nearing end of life
Remove sensor from housing. Using a volt-meter set
to uA output; apply the (+) lead to the outer ring of
the sensor PCB and the (-) lead to the center circle
to obtain the sensor’s output in air. Contact factory
with result.
Sensors without PCB use mV setting.
Calibrate the analyzer (calibrate at pressure and
temperature of sample)
Clean contacts with alcohol (minimize exposure time
of MS sensor to ambient air to extent possible)
Replace sensor and return sensor to the factory for
warranty determination
Upper section of sensor housing: Clean contacts with
alcohol, flow sample or zero gas for 2-3 hours to
flush sample system and sensor housing
Sensor: Replace if leaking and return it to the factory
for warranty determination
Wipe with alcohol and lint free towel or flow sample
or zero gas for 2-3 hours to flush
Replace GPR/PSR sensor with XLT sensor when CO
2
or acid gases are present
Consult factory
Replace sensor and install scrubber
Replace sensor, obtain authorized service
Replace sensor
Erratic O
2
reading
or
Negative O
2
reading
or
No O
2
readin
g
possibly
accompanied by
electrolyte leakage
Pressurizing the sensor by flowing gas to
the sensor with: the vent restricted
and suddenly removing the
restriction draws a vacuum on the sensor
or
partially opening the valves upstream of
the analyzer when using a pump
downstream of the analyzer to draw
sample from a process at atmospheric
pressure or a slight vacuum.
A pressurized sensor may not leak but
still produce negative readings.
A premature adjustment of the ZERO
OFFSET potentiometer is a common
problem
Zero the analyzer. If not successful replace the
sensor
Avoid drawing a vacuum on the sensor.
Placing a vacuum on the sensor in excess 4” of
water column is strongly discouraged.
From MAIN MENU select DEFAULT ZERO
Vue de la page 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32

Commentaires sur ces manuels

Pas de commentaire