Installing the Oxygen Sensor
The GPR-1500-A Oxygen Transmitter is equipped with a SS sensor housing. This housing offers ease of
replacement of sensor and at the same time prevents any air leakage into the system. The two sections of
the sensor are held together by a metal clamp secured in place by easily accessed bolt. The integrity of the
sensor housing has been tested at the factory prior to shipment and is fully operational from the shipping
container.
Caution: All transmitters must be calibrated once the
installation has been completed and periodically thereafter
as described below. Following the initial installation and
calibration, allow the transmitters to stabilize for 12-24
hours and re-calibrate the transmitter with a certified span gas.
Caution: DO NOT dissect the oxygen sensor. The
sensor contains a corrosive liquid electrolyte that could
be harmful if touched or ingested, refer to the Material
Safety Data Sheet contained in the Owner’s Manual appendix. Avoid
contact with any liquid or crystal type powder in or around the sensor
or sensor housing, as either could be a form of electrolyte. Leaking
sensors should be disposed off in a manner similar to that of a
common battery in accordance with local regulations.
Avoid electrostatic discharge – Clean all surfaces with a
damp cloth only.
Procedure
1. Loosen the bolt at the bottom of the sensor housing by using 5/16 ranch provided.
2. Twist the upper section of the housing 90 degree and pull it up until it clears the bottom section of the
sensor housing.
3. Remove the old sensor (if previously installed) from the sensor housing
4. Remove the oxygen sensor from the bag and remove the two red shorting taps from the two ring gold
color contact plate of the sensor.
Remove the two red ribbons
from sensor PCB