Advanced Instruments Inc.
2855 Metropolitan Place, Pomona, California 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, e-mail: [email protected]
GPR-2900
Oxygen Analyzer
Owner’s Manual
Vue de la page 0
1 2 3 4 5 6 ... 36 37

Résumé du contenu

Page 1 - Oxygen Analyzer

Advanced Instruments Inc. 2855 Metropolitan Place, Pomona, California 91767 USA ♦ Tel: 909-392-6900, Fax: 909-392-3665, e-mail: [email protected]

Page 2 - 1 Introduction

Advanced Instruments Inc. 10 Air calibrate the analyzer (with the exception of Oxygen Purity Analyzers intended to measure elevated oxygen l

Page 3

Advanced Instruments Inc. 11 Recommendations to avoid erroneous oxygen readings and damaging the sensor: ¾ Do not place your finger over th

Page 4 - 3 Safety Guidelines

Advanced Instruments Inc. 12 Gas Connections: The GPR-2900 with its standard flow through configuration is designed for positive pressure sa

Page 5

Advanced Instruments Inc. 13 4. To connect to an active relay or “fail safe”, connect the live cable to the common terminal C and the secon

Page 6 - 5 Operations

Advanced Instruments Inc. 14 Installing the Oxygen Sensor The GPR-2900 Oxygen Analyzer is equipped with an external oxygen sensor. They have

Page 7

Advanced Instruments Inc. 15 Establishing Power to the Electronics: Once the power to the electronics is established, the digital display re

Page 8

Advanced Instruments Inc. 16 Note: For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the fu

Page 9

Advanced Instruments Inc. 17 9. Once the OVER RANGE warning appears the user must advance the analyzer to the next higher range via the men

Page 10 - Advanced Instruments Inc

Advanced Instruments Inc. 18 10. Note: The PERCENT alarm value is entered with one decimal, the PPM alarm value is entered as an integer.

Page 11

Advanced Instruments Inc. 19 Set Alarm Delay: Once the values for ALARM 1 and ALARM 2 have been entered, the user may elect to delay the a

Page 12

Advanced Instruments Inc. 2 Table of Contents Introduction 1 Quality Control Certification 2 Safety 3 Features & Specificati

Page 13

Advanced Instruments Inc. 20 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATION CONFIG ALARMS BYPASS ALARMS >>> MAIN MENU SET A

Page 14

Advanced Instruments Inc. 21 Bypass Alarms: This feature, separate from CONFIG ALARMS above, enables the user to disable the local audible

Page 15

Advanced Instruments Inc. 22 Zero Calibration In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample g

Page 16 - 3.3 %

Advanced Instruments Inc. 23 0.000 PPM ZERO CALIBRTION ENTER TO CALIBRATE MENU TO ABORT 8. Press the ENTER key to calibrate or MENU k

Page 17

Advanced Instruments Inc. 24 Output Zero: In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This

Page 18

Advanced Instruments Inc. 25 Span Calibration Maximum drift from calibration temperature is approximately 0.11% of reading per °C. The analy

Page 19

Advanced Instruments Inc. 26 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATION CONFIG ALARMS BYPASS ALARMS 3.3% AUTO SAMPLING 10 % RANGE

Page 20

Advanced Instruments Inc. 27 23. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key. 24. Allow approxi

Page 21

Advanced Instruments Inc. 28 Output Span: In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This

Page 22

Advanced Instruments Inc. 29 Sampling GPR-2900 Oxygen Analyzer requires positive pressure to flow the sample gas by the sensor to measure th

Page 23

Advanced Instruments Inc. 3 2 Quality Control Certification Date: Customer: Ord

Page 24

Advanced Instruments Inc. 30 6 Maintenance There are no moving parts in the analyzer given the modular nature of the electronics and sensor

Page 25

Advanced Instruments Inc. 31 8 Troubleshooting Symptom Possible Cause Recommended Action Reading does not reflect expected values Se

Page 26

Advanced Instruments Inc. 32 9 Warranty The design and manufacture of GPR Series oxygen analyzers, monitors and oxygen sensors are perform

Page 27

Advanced Instruments Inc. 33 10 MSDS Material Safety Data Sheet Product Identification Product Name Oxygen Sensor Series - PSR, GPR, AII,

Page 28

Advanced Instruments Inc. 34 Spill or Leak Steps if material is released Sensor is packaged in a sealed plastic bag, check the sensor ins

Page 29

Advanced Instruments Inc. 35 Appendix B Correlating Readings - LCD Display and 4-20mA Output In rare instances the 4-20mA signal output may

Page 30 - 7 Spare Parts

Advanced Instruments Inc. 36 100.0 OUTPUT ZERO OFFSET PRESS UP OR DOWN TO CHANGE VALUE ENTER TO SAVE MENU TO RETURN 14. Enter th

Page 31 - 8 Troubleshooting

Advanced Instruments Inc. 37 27. Enter the calculated adjustment value, refer to example described above. Note: Once the initial adjustment

Page 32 - 9 Warranty

Advanced Instruments Inc. 4 3 Safety Guidelines General This section summarizes the essential precautions applicable to the GPR-2900 Oxyg

Page 33

Advanced Instruments Inc. 5 Maintenance Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for t

Page 34

Advanced Instruments Inc. 6 5 Operations Principle of Operation The GPR-2900 Oxygen Analyzer incorporates a variety of percentage range ad

Page 35 - Appendix B

Advanced Instruments Inc. 7 Calibration & Accuracy Overview Single Point Calibration: As previously described the galvanic oxygen

Page 36

Advanced Instruments Inc. 8 Recommendation 1: Zero calibration, see Determining True Zero Offset below, is recommended only for online anal

Page 37

Advanced Instruments Inc. 9 Installation Considerations Gas Sample Stream: Ensure the gas stream composition of the application is consisten