Table of Contents
1 Introduction 1
1.1 Indications for Use
1.2 Intended Use
1.3 Device Description
1
2
2
2 Quality Control Certification 3
3 Safety Warnings 4
4 Start-up 6
4.1 Contents of Shipping Container
4.2 Controls
4.3 Start-Up Test
4.4 Calibration
4.5 Mounting
6
6
6
7
9
5 Operation 10
5.1 Principle of Operation
5.2 Application Considerations
5.3 Calibration
5.4 Sampling
10
11
12
12
6 Maintenance 14
7 Troubleshooting 18
8 Specifications 19
9 Warranty 22
10 Material Safety Data Sheet (MSDS) 23
6.1 Serviceability
6.2 Battery Replacement
6.3 Oxygen Sensor Replacement
6.4 Reassembly
6.5 Remote Oxygen Sensor Replacement
14
14
15
16
17
8.1 Spare Parts & Optional Accessories 20
10.1 Disposal 23
i
1 Introduction
Congratulations on your purchase, these Instructions for Use describe the pre-
cautions, set-up, operation, maintenance and specifications of the Palm O2
Oxygen Analyzer.
This symbol means CAUTION – Failure to read and comply with the
Instructions for Use could damage the device and possibly jeopardize
the well being of the user.
Note: Advanced Instruments Inc. cannot warrant any damage resulting from
the misuse, unauthorized repair or improper maintenance of the device.
1.1 Indications for Use
The Palm O2 Oxygen Analyzer is intended to measure and display the concen-
tration of oxygen for personnel safety, area monitoring, O2 deficiency in con-
fined spaces, checking breathing air tanks and confirming the O2 levels prior to
welding.
Users must read the following statements as they are essential to re-
ducing the risk of use error due to ergonomic features of the device or
the environment in which the device is intended to be used.
The device has been designed and manufactured in such a way that when used
under the conditions and for the purposes intended, they will not compromise
the safety of the users or other persons.
Conformity with essential requirements has been demonstrated by verifying the
performance of the device under normal conditions, bench testing and deter-
mining that undesirable malfunctions constitute minimal risk to users.
Do not sterilize, autoclave, liquid sterilize, immerse in any liquid or expose the
device or accessories to steam, ethylene oxide or radiation sterilization.
The device is intended to be re-usable. Should the device or accessories come
in contact with patient bodily fluids, either dispose of the device or clean with a
soft cloth dampened with 70% isopropyl alcohol solution in water and allow the
components to air-dry before re-use .
Do not operate the analyzer near equipment capable of emitting high levels of
electromagnetic radiation as the reading may become unstable.