Analytical-industries GPR-1500 A Series Trace PPM Oxygen Analyzer Manuel d'utilisateur Page 37

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 51
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 36
Advanced Instruments, Inc
37
or
No O
2
reading
accompanied by
electrolyte leakage
suddenly removing the
restriction draws a vacuum
on the sensor
or
partially opened valves
upstream of the transmitter
when using a pump
downstream of the
transmitter to draw sample
from a process at
atmospheric pressure or
under a slight vacuum.
Placing a vacuum on the
sensor in excess 10” of water
column is strongly
discouraged.
A premature adjustment of
the ZERO OFFSET
potentiometer is a common
problem
Avoid drawing a vacuum on the sensor, a
pressurized sensor may not leak but still
produce negative readings.
From MAIN MENU select DEFAULT ZERO
Vue de la page 36