Mounting the Analyzer
The standard GPR-1600 is designed to be panel mounted and requires a cutout that accommodates the enclosure and 4
mounting bolts. The design also lends itself to 19 rack mounting with an optional bezel or wall mount enclosures as illustrated
below.
Procedure:
1. The GPR-1600 is designed for panel mounting directly to any flat vertical surface, wall or bulkhead plate with the
appropriate cut out and four ¼” diameter holes for insertion of the mounting studs located on the back side of the front
panel.
2. When mounting the analyzer position it approximately 5 feet off the floor for viewing purposes and allow sufficient room for
access to the terminal connections at the rear of the enclosure.
3. Note: The proximity of the analyzer to the sample point and use of optional sample conditioning components have an
impact on sample lag time.
Mounting GPR-1600: