Analytical-industries GPR-1600 Series Trace PPB Oxygen Analyzer Manuel d'utilisateur Page 44

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 55
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 43
Advanced Instruments Inc.
44
CALIBRATION
COMPLETE
Following a successful calibration, the system returns to theSample mode and it will be necessary to reconnect the sample
gas line as follows:
Span Gas Calibration: Assure there are no restrictions in vent line.
Regulate the pressure between 5-30 psig and set the flow rate to 2 SCFH for Fuel Cell sensors.
If the analyzer is equipped with a SAMPLE/BYPASS valve, place it in the BYPASS position.
Disconnect the span gas line and install the sample gas line.
Keep the SAMPLE/BYPASS valve in the BYPASS position.
Allow the sample gas to flow for 1-2 minutes to purge the gas lines inside the analyzer.
Place the SAMPLE/BYPASS valve in the SAMPLE position.
If the analyzer is not equipped with a SAMPLE/BYPASS valve:
Allow the sample gas to flow for 1-2 minutes to purge the air trapped in the sample gas lines.
Disconnect the span gas line and install the purged sample gas line.
Air Calibration:
The sample gas should still be flowing, if necessary open the FLOW valve.
Reassemble the sensor housing.
Place the sensor face down in the bottom section of the sensor housing.
Ensure the o-ring is in place.
Install the upper section by gently pushing it
straight down.
Ensure the PCB contacts the two gold pins.
Rotate the upper section of the sensor housing
90º to engage the clamp.
Using the 5/16 wrench supplied tighten the clamp
bolt located in the center of the bracket attached
to bottom section with the elbow fittings to
secure the sensor housing.
Close the front door of the analyzer.
Vue de la page 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 54 55

Commentaires sur ces manuels

Pas de commentaire