Analytical-industries GPR-28 Oxygen Analyzer Manuel d'utilisateur Page 25

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 39
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 24
Advanced Instruments Inc.
25
The above times assume the introduction of a purge gas, the lower of the available zero or
sample (if known and constant) gas, with a ppm level O2 concentration less than 0.5 ppm is
introduced to the analyzer following span calibration to purge (accelerate the reaction of)
the O2 that has dissolved into the electrolyte inside the sensor. If zero gas is not available,
substitute the sample gas and expect slightly longer recovery time.
Span Gas Selection:
The O2 concentration of a span gas should approximate 70-90% of the full scale range
dictated by the span gas, e.g. 80 ppm O2 on the 0-100 ppm range. For optimum accuracy,
the full scale range dictated by the span gas should be at least one range higher than the
intended analysis range. Both of the aforementioned recommendations reduce the error
induced by the tolerance of the electronic components; also, span gases with higher O2
concentrations are more reliable and less expensive.
Conversely, if the recommended span gas is not available and air calibration is not an
option, a span gas of the same full scale range and near the anticipated analysis level
(approximately 10% of full scale) is acceptable with the understanding the accuracy will
suffer slightly.
Use of span gas near 10% of the same full scale range for measurements at the higher end
of the range has the effect of “expanding the error” by moving upscale as illustrated by
Graph A and Example 1 in the Accuracy section above and is not recommended. Of course
the user can always elect at his discretion to accept an accuracy error of +
2-3% of full scale
range if no other span gas is available.
Type of Sensor:
Pico-Ion MS Oxygen Sensors: Analyzers utilizing these sensors require span gas(es) and
cannot be calibrated with air (20.9% or 209,000 ppm O2) due to their high signal output.
Calibrate with a span gas containing less than 1000 ppm O2; a span gas containing 70-90
ppm O2 is recommended.
Galvanic Fuel Cell Sensors: Analyzers utilizing these sensors can be calibrated with ambient
air (20.9% or 209,000 ppm O2) and have been proven capable of precise analysis below
0.5 ppm (500 ppb).
Note: As described above, these oxygen sensors are capable of one point calibration.
Span Gas vs. Air
(Fuel Cell Sensors only):
Span gas calibration: Recommended for calibration of in-service analyzers, if the priority is
the fastest online recovery time.
Air calibration: Recommended for calibration an analyzer if a new oxygen sensor is being
installed or if the availability of span gas, the cost of span gas or the accuracy of a span gas
is an issue. An air calibrated analyzer can be used to reliably verify a “certified” span gas,
which has frequently been found to be inaccurate. For best results, select a recognized
name supplier.
Note: Galvanic Fuel Cell sensors can be calibrated (using air) and brought online (using
sample gas) without “purchased calibration gases” and without sacrificing accuracy -
provided the analyzer is given enough time to come down to the O2 concentration of the
sample gas.
Vue de la page 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 38 39

Commentaires sur ces manuels

Pas de commentaire